EBI 12-TPX9X Pressure Logger
|
Temperature / Pressure Data Loggers EBI 12 series
General specifications for all EBI 12-TP series |
|
Operating temperature (pressure logger) |
0 °C ... +150 °C (+32 °F ... +302 °F) |
Temperature |
|
Accuracy |
±0.1 °C (0 °C ... +120 °C) ±0.05 °C (+120 °C ... +140 °C) ±0.1 °C (+140 °C … +150 °C) |
Resolution |
0.01 °C |
Sensor |
Pt 1000, Class A |
Pressure |
|
Accuracy |
±10 mbar (50 mbar … 150 mbar) ±10 mbar (2,050 mbar … 2,250 mbar) ±10 mbar (3,000 mbar … 3,250 mbar) ±15 mbar (for the remaining measurement range) |
Resolution |
1 mbar |
Sensor |
Piezoresistive pressure sensor (temperature compensated) |
Data memory |
Max. 100,000 measurements (total) |
Sampling rate |
250 ms ... 24 |
Measurement mode |
• Endless measurement • Measure upon start time • Start immediately until end of memory • Start- / Stop measurement |
Storage temperature |
0 °C ... +125 °C (+32 °F ... +257 °F) |
Maximum operating pressure |
Pressure logger: 7 bar (abs) |
Minimum operating pressure |
Logger with flexible cable probe: 20 mbar (abs) |
Time accuracy (24h) |
< 5 sec |
Battery |
Lithium cell, 3.6 V, user replaceable |
Dimensions (Ø x H) |
48 mm x 32 mm* |
Weight |
Approximately 120 g * |
Housing material |
316 L Stainless steel / PEEK |
Protection class |
IP 68 |
Certificate |
Factory calibration certificate |
Product lines
|
Red line |
Blue line |
Short Description |
Temperature data logger with external probe less than 2 mm diameter Data logger with additional units Special process logger |
All other temperature data logger |
Radio mode |
Loggers can transfer data during the measurement process |
Not possible |
Total memory capacity |
100,000 values |
27,000 values |
Click here to download brochure
Untuk info lengkap mengenai Harga dan Stok silahkan kontak kami via Online Chat atau Contact
EBI 12-TPX9X Precision Pressure Logger
high precision pressure measurement down to 0.1 mbar (Red Line)
In the process of H2O2 Sterilizers a very precise pressure measurement is necessary.
The system works withpressures down to 1 mbar abs. A Highly accurate pressure data logger is necessary caused by the very lowpressure nearly by vacuum.
• 1 internal or 1 external temperature sensor, Ø 1.95 mm
• 1 internal pressure sensor
*Not for use in steam sterilizer
Specification |
|
Temperature |
|
Measurement range |
0 °C ... +85 °C (+32 °F ... +185 °F) |
Accuracy |
±0.1 °C |
Resolution |
0.01 °C |
Sensor |
Pt 1000 |
Pressure |
|
Measurement range |
0.1 … 1,050 mbar (0.1 ... 788 Torr) |
Accuracy |
± 0.25 mbar (0.1 mbar … 50 mbar) ± 5% of measured value (50 mbar … 100 mbar) ± 1% FS (100 mbar … 1,050 mbar) |
Resolution |
0.1 mbar |
Sensor |
Piezoresistive pressure sensor |
Data memory |
TP190: 50,000 measurement values TP290: 2 x 33,400 measurement values |
Measurement cycle |
250 ms ... 24 hours |
Measurement mode |
• Endless measurement immediately • Start / Stop measurement • Measure immediately until end of memory |
Operating temperature |
0 °C … +85 °C |
Max. operating pressure |
2,000 mbar (abs.) |
Battery |
Lithium cell, 3.6 V, user replaceable |
Dimensions (D x H) |
48 mm x 35 mm |
Housing material |
316 L Stainless steel / PEEK |
Protection class |
IP 68 |
Factory calibration certificate |
0 °C and +60 °C <=0.5 mbar, 50 mbar at +25 °C and <=0.5 mbar, 10 mbar, 50 mbar at +60 °C |
Ordering Information
Type |
Description |
Part No. |
EBI 12-TP190 |
Internal temp. sensor |
1340-6665 |
EBI 12-TP290 |
Needle Length = 40 mm |
1340-6666 |