EBI 12-TC230 Conductivity Data Logger
Cat. No. : 1340-6667
Brand
Ebro
Berat
1 Kg
Stok
indent
Deskripsi
Spesifikasi Teknis

Temperature / Pressure Data Loggers EBI 12 series

General specifications for all EBI 12-TP series

Operating temperature

(pressure logger)

0 °C ... +150 °C (+32 °F ... +302 °F)

Temperature

 

Accuracy

±0.1 °C (0 °C ... +120 °C)

±0.05 °C (+120 °C ... +140 °C)

±0.1 °C (+140 °C … +150 °C)

Resolution

0.01 °C

Sensor

Pt 1000, Class A

Pressure

 

Accuracy

±10 mbar (50 mbar … 150 mbar)

±10 mbar (2,050 mbar … 2,250 mbar)

±10 mbar (3,000 mbar … 3,250 mbar)

±15 mbar (for the remaining measurement range)

Resolution

1 mbar

Sensor

Piezoresistive pressure sensor (temperature compensated)

Data memory

Max. 100,000 measurements (total)

Sampling rate

250 ms ... 24

Measurement mode

• Endless measurement

• Measure upon start time

• Start immediately until end of memory

• Start- / Stop measurement

Storage temperature

0 °C ... +125 °C (+32 °F ... +257 °F)

Maximum operating pressure

Pressure logger: 7 bar (abs)

Minimum operating pressure

Logger with flexible cable probe: 20 mbar (abs)

Time accuracy (24h)

< 5 sec

Battery

Lithium cell, 3.6 V, user replaceable

Dimensions (Ø x H)

48 mm x 32 mm*

Weight

Approximately 120 g *

Housing material

316 L Stainless steel / PEEK

Protection class

IP 68

Certificate

Factory calibration certificate

 

 

Product lines

 

Red line

Blue line

Short Description

Temperature data logger with external

probe less than 2 mm diameter

Data logger with additional units

Special process logger

All other temperature data logger

Radio mode

Loggers can transfer data during the

measurement process

Not possible

Total memory capacity

100,000 values

27,000 values

 

Click here to download brochure

Untuk info lengkap mengenai Harga dan Stok silahkan kontak kami via Online Chat atau Contact

EBI 12-TC230 Temperature / Conductivity Data Logger 1340-6667

Conductivity measurement (Red Line)

 

In processes like e.g. washer disinfectors, the measurement of conductivity in the last dishwater is required.This is reasonably done in the running process without any interruption.

 

• 1 external temperature probe

• 1 external conductivity probe

 

 

Specification

Temperature

 

Measurement range

0 °C ... +125 °C (+32 °F ... +257 °F)

Accuracy

± 0.2 °C

Resolution

0.01 °C

Sensor

Pt 1000

Conductivity

 

Measurement range

1 … 2,000 μS/cm

Accuracy

± 0.5 μS (1 μS/cm .. 20 μS/cm)

± 1.0 μS (20.1 μS/cm .. 100 μS/cm)

± 2.5% of measured value (100.1μS/cm ... 1500μS/cm)

± 5% from the measurement range (remaining range)

Resolution

0.1 μS/cm

Sensor

Conductivity electrode

Data memory

2 x 50,000 measurement values

Measurement mode

• Endless measurement immediately

• Start / Stop measurement

• Measure immediately until end of memory

Storage temperature

0 °C ... +125 °C (+32 °F ... +257 °F)

Max. operating pressure (burst pressure)

2.5 bar (abs.)

Battery

Lithium cell, user replaceable

Dimensions (D x H)

48 mm x 130 mm

Housing material

316 L Stainless steel / PEEK

Protection class

IP 68

Factory calibration certificate

0 °C, +60 °C and +121 °C

5 μS/cm, 100 μS/cm, 1413 μS/cm at +25 °C

Radio mode

Not possible

 

Produk Lainnya
GLOBAL SCIENTIFIC INTERNATIONAL
Sentra Niaga Puri Indah Blok T3-6
Kembangan Selatan - Jakarta Barat
DKI Jakarta 11610 - Indonesia
www.gsi.id
Customer Support
Shipping and Payment Method