EBI 12-T22X Temperature Data Logger
|
General specifications for EBI 12 series
Operating temperature: logger |
-90 °C … +150 °C (-130 °F ... 302 °F) |
Operating temperature: radio operation |
-30 °C … +150 °C (-22 °F ... 302 °F) |
Temperature accuracy |
±1.5 °C (-200 °C … -85 °C) ±0.5 °C (-85 °C … -40 °C) ±0.2 °C (-40 °C … 0 °C) ±0.1 °C (0 °C ... +120 °C) ±0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C) ± 0.5 °C (+150 °C … +250 °C) ± 0.8 °C (+250 °C … +400 °C) |
Time accuracy |
< 5 sec (24 h) |
Temperature resolution |
0.01 °C |
Sensor |
Pt 1000, Class A |
Interval |
250 ms ... 24 hours |
Measurement mode |
• Endless measurement • Start / stop measurement • Measure upon start temperature • Start immediately until end of memory |
Storage temperature |
0 °C … +125 °C |
Maximum operating pressure |
Temperature logger 10 bar (abs) |
Minimum operating pressure |
Logger with flexible cable probe 20 mbar (abs) |
Battery |
Lithium cell, 3.6 V, user replaceable |
Dimensions (Ø x H) |
48 mm x 24 mm |
Weight |
Approximately 110 g |
Housing material |
316 L Stainless steel / PEEK |
Protection class |
IP 68 |
Certificate |
Factory calibration certificate |
Product lines
|
Red line |
Blue line |
Short Description |
Temperature data logger with external probe less than 2 mm diameter Data logger with additional units Special process logger |
All other temperature data logger |
Radio mode |
Loggers can transfer data during the measurement process |
Not possible |
Total memory capacity |
100,000 values |
27,000 values |
Click here to download brochure
Untuk info lengkap mengenai Harga dan Stok silahkan kontak kami via Online Chat atau Contact
EBI 12-T22X Temperature Data Logger
bendable metal probe (Red Line)
• 1 external temperature probe, Ø 1.5 mm
• Radial arranged
Application
• Measurement in drain of sterilizers
• In oven (protected by Thermal-Isolation-Box)
• EtO sterilization (ATEX type)
Specification
Measurement range (Use in EX Area) |
EBI 12-T220 : -200 °C ... +200 °C (-328 °F ...+392 °F) EBI 12-T220-EX : -40 °C ... +85 °C (-40 °F ...+185 °F) EBI 12-T221 : -200 °C ... +400 °C (-328 °F ... +752 °F) |
Accuracy |
EBI 12-T220 : ± 1,5 °C (-200 °C … -85 °C) ± 0,5°C (-85 °C … -40 °C) ± 0,2°C (-40 °C … 0 °C) ± 0,1 °C (0 °C ... +120 °C) ± 0,05 °C (+120 °C … +140 °C) ± 0,1 °C (+140 °C … +150 °C) EBI 12-T220-EX : ± 0,2°C (-40 °C … 0 °C) ± 0,1 °C (0 °C ... +120 °C) ± 0,05 °C (+120 °C … +140 °C) ± 0,1 °C (+140 °C … +150 °C) EBI 12-T221 : ± 1,5 °C (-200 °C … -85 °C) ± 0,5°C (-85 °C … -40 °C) ± 0,2°C (-40 °C … 0 °C) ± 0,1 °C (0 °C ... +120 °C) ± 0,05 °C (+120 °C … +140 °C) ± 0,1 °C (+140 °C … +150 °C) ± 0,5 °C (+150 °C …+ 250 °C) ± 0,8 °C (+250 °C … +400 °C) |
Data memory |
100,000 measurement values |
Factory calibration certificate |
-20 °C, 0 °C, +60 °C, +134 °C and +250 °C if the probe length is 500 mm |
Sensor |
Pt 1,000, class A |
Sample Rate |
250 msec ... 24 h |
IP Rating |
IP 68 |
Battery |
Lithium battery (3,6 V), user replaceable |
Dimensions |
(Ø x H) 48 mm x 24 mm |
Ordering Information
Type |
Description |
Cat. No. |
EBI 12-T220 |
Needle length = 250 mm |
1340-6604 |
EBI 12-T220-EX |
Needle length = 250 mm |
1340-6604-EX |
EBI 12-T221 |
Needle length = 500 mm |
1340-6605 |