BR2 Bilirubin Stat-Analyzer
BR2 Bilirubin Stat-Analyzer
Brand
Advanced Instruments
Weight
5.5 Kg
Stock
indent
Description
Technical Specification

The Advanced® BR2 Bilirubin Stat-Analyzer™ system provides fast, accurate results for both total and direct bilirubin levels in neonates requiring only a 30µL serum or plasma sample. It determines total bilirubin by spectrophotometric analysis and direct bilirubin using a modified version of the proven Malloy-Evelyn method. Because of its compact, portable design and ease of use, it is ideally suited for pediatric physician offices, satellite labs, hospital stat labs, or pediatric research laboratories.

- Sample Size : 30 μL

- Sample Type : Serum or Plasma

- Sample Capacity : Single sample, disposable cuvettes

- Test Time :

  * Total : Instantaneous

  * Direct : 2 minutes

- Units : mg/dL Bilirubin (also available in μmol/L) S.I. units for export only; 1 mg/dL bilirubin = 17.1 μmol/L)

- Range : 0 to 30 mg/dL

- Linearity : Within ±0.2 mg/dL from straight-line from 0 to 30 mg/dL

- Resolution : ±0.1 mg/dL

- Repeatability : Within ±0.2 mg/dL over entire range

- Drift : Zero: < 0.2 mg/dl after 30-minute warm up

- Calibration: < 0.2 mg/dL 24 hr after warm up

- Hemolysis : No interference from HbO 2 levels up to 600 mg/dL

- Turbidity : Negligible interference from turbidity associated with triglyceride (TGL) levels up to 500 mg/dL

- Operating Conditions : Temperature 15°C to 32°C (59°F to 90°F) ; 5% to 80% relative humidity (non condensing)

- Storage Temperature : -40°C to +70°C (-40°F to +160°F)

- Electrical Voltage : 100 to 120 VAC (60 Hz) or 200 to 240 VAC (50Hz)

- Power Consumption : 15 W