BR2 Bilirubin Stat-Analyzer
BR2 Bilirubin Stat-Analyzer
|
The Advanced® BR2 Bilirubin Stat-Analyzer™ system provides fast, accurate results for both total and direct bilirubin levels in neonates requiring only a 30µL serum or plasma sample. It determines total bilirubin by spectrophotometric analysis and direct bilirubin using a modified version of the proven Malloy-Evelyn method. Because of its compact, portable design and ease of use, it is ideally suited for pediatric physician offices, satellite labs, hospital stat labs, or pediatric research laboratories.
- Sample Size : 30 μL
- Sample Type : Serum or Plasma
- Sample Capacity : Single sample, disposable cuvettes
- Test Time :
* Total : Instantaneous
* Direct : 2 minutes
- Units : mg/dL Bilirubin (also available in μmol/L) S.I. units for export only; 1 mg/dL bilirubin = 17.1 μmol/L)
- Range : 0 to 30 mg/dL
- Linearity : Within ±0.2 mg/dL from straight-line from 0 to 30 mg/dL
- Resolution : ±0.1 mg/dL
- Repeatability : Within ±0.2 mg/dL over entire range
- Drift : Zero: < 0.2 mg/dl after 30-minute warm up
- Calibration: < 0.2 mg/dL 24 hr after warm up
- Hemolysis : No interference from HbO 2 levels up to 600 mg/dL
- Turbidity : Negligible interference from turbidity associated with triglyceride (TGL) levels up to 500 mg/dL
- Operating Conditions : Temperature 15°C to 32°C (59°F to 90°F) ; 5% to 80% relative humidity (non condensing)
- Storage Temperature : -40°C to +70°C (-40°F to +160°F)
- Electrical Voltage : 100 to 120 VAC (60 Hz) or 200 to 240 VAC (50Hz)
- Power Consumption : 15 W